  OT-2-3 Optical Tester
OT-2-3 Optical Tester
OT-2-3 Optical Tester
The OT-2-3 is designed for measurement of optical power and attenuation in optical fibers and fiber optical components with high accuracy.

It can be used as a standard instrument for calibration of optical testers, power meters and laser sources.

The OT-2-3 has a power meter as a head, three high stability laser sources and a built-in attenuator. Laser power of the LD is regulated by laser current, which can be changed by the PC commands.

A 1 mm-InGaAs photodiode is used for optical power measurement.

The needed range and measurement accuracy are provided by the automatic switch of the optical receiver gain.

The OT-2-3 control and data processing are provided by a PC under Windows 95/98/NT (software screenshot).
  Specifications
Laser Source
Wavelength, nm
850±5
1310±5, 1550±5
Power, dBm
2
3
Stability, dB:
- 15 min
0.005
- 1 h
0.02
Power meter
Power range, dBm
+3..-70
+3..-80
Uncertainty, %
4 (+3..-70) dBm
3 (+3..-60) dBm
3 (+3..-60) dBm
4 (+3..-70) dBm
Linearity, %
1.2 (+3..-70) dBm
0.8 (+3..-60) dBm
1.2 (+3..-60) dBm
0.8 (+3..-70) dBm
Uncertainty in the spectral range 780…1630 nm, %
5
Resolution, dB
0.001
Built-in attenuator range, dB
0..40
Power Supply, V
~220
Dimensions, mm
293x255x60
Weight, kg
3